Transcript: Automated Blood Alcohol Analysis With Gas Chromatography
1960s
These captions and transcript were generated by a computer and may contain errors. If there are significant errors that should be corrected, please let us know by emailing digital@sciencehistory.org.
00:00:00 This film is dedicated to Dr. G. Machata and his team at Perkin-Elmer.
00:00:30 Blood alcohol analysis with gas chromatography seems to be the most modern procedure.
00:00:49 Besides the determination of ethanol, other volatile compounds can be easily detected.
00:00:54 Automation of this procedure was necessary to increase the speed as well as the precision of the analysis.
00:01:01 The best way was to apply headspace analysis in combination with a fully automated sample injection system.
00:01:08 The first equipment of this type was developed by Perkin-Elmer, Germany, in cooperation with the author of this film.
00:01:15 It has been under continuous operation since April 1968 at the Institute of Forensic Medicine of the University of Vienna.
00:01:24 The so-called Multifract F-40 consists of a temperature-controlled sample changer designed as a turntable,
00:01:31 the usual gas chromatograph with a special dosing system, the programmer for controlling the automatic cycle,
00:01:37 the pressure regulating unit, and the recorder.
00:01:41 To increase the accuracy of the method, an internal standard is added to the sample,
00:01:45 and the quotient of the individual peak heights is determined.
00:01:49 The absolute alcohol content may be read on a calibration curve.
00:01:53 The difference of the vapor pressures in serum or blood is accounted for by calculation.
00:02:04 Tertiary butanol is used as an internal standard.
00:02:08 The vapor pressure curve of this substance runs parallel with that of ethanol over a wide range of temperatures.
00:02:29 This scheme shows the construction of the new electro-pneumatic dosing system.
00:02:33 The carrier gas first flows through a reducing valve and then through a solenoid valve into the column.
00:02:39 A small part of the carrier gas is used for continuous cleaning of the dosing capillary.
00:02:44 Afterwards, it passes through the movable sealing cylinder and then through the cleaning line and a valve to the exterior.
00:02:52 A silicone rubber membrane closes the sealing cylinder,
00:02:56 and a solenoid valve in the cleaning line regulates the pressure in the sealing cylinder.
00:03:09 After pushing the start button, the turntable goes to the next position and is lifted to the upper position.
00:03:19 The dosing capillary pierces the seals of the cylinder and of the sample bottles.
00:03:23 The carrier gas flows into the bottle until a corresponding pressure is built up.
00:03:33 By closing the solenoid valve of the carrier gas,
00:03:36 a small amount of the sample is transferred into the gas chromatographic column.
00:03:41 The flow of sample vapor into the column is interrupted by opening the solenoid valve again.
00:03:53 The turntable remains in the upper position for a short time at the beginning of the gas chromatographic separation.
00:04:00 This avoids disturbing the carrier gas flow.
00:04:12 The turntable then returns to the lower position and is now in state.
00:04:23 At the end of the analysis, the turntable takes up the middle position.
00:04:27 The end of the dosing capillary is now between the seals of the dosing cylinder and the sample bottle.
00:04:33 A strong flow of carrier gas frees the capillary from the remaining sample vapors.
00:04:42 The turntable again moves to the lower position and goes on to the next sample.
00:04:54 The samples and the internal standard are placed in special sample bottles by normal or automatic pipettes.
00:05:24 The bottles are closed with rubber septums and secured with aluminum caps.
00:05:55 The bottles containing the samples are inserted into the thermostating turntable, heated to 60 degrees centigrade.
00:06:12 The different steps of analysis will be selected on the programmer.
00:06:17 After conditioning the gas chromatograph, the analysis is started.
00:06:24 The automatic sequence is marking the sample on the diagram,
00:06:31 pressure increasing in the sample bottle, sample withdrawal,
00:06:42 analysis,
00:07:08 cleaning of the dosing capillary,
00:07:15 changing to the next sample,
00:07:22 marking the sample.
00:07:27 The parameters of analysis can be varied extensively,
00:07:30 and manual injection is possible through a separate injection port.
00:07:34 A zero adjustment device automatically corrects the baseline.
00:07:41 The accuracy of the analysis is very high.
00:07:44 Evaluation by electronic integrator and further by computer is possible.
00:07:52 The most frequently used stationary phase is Carbowax 1500.
00:07:56 For special separations, Halkomid or Poropak columns are used.
00:08:01 The multifract F40 is used in our laboratory for the routine determination of blood alcohol,
00:08:07 as well as for the clinical estimation of acetone, ether, and fluotan in blood during narcosis.
00:08:13 Two hours instrument time are required for 30 samples of blood,
00:08:17 and one more hour for the preparation of the samples and for the calculation of the results.
00:08:31 www.ottobock.com